362 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
viVPrintf(AgtE4980A/AL, "LIST:STIM:DATA 1E3, 1, 2E3, 1.1, 3E3,
0.9" + vbLf, 0)
Equivalent key No equivalent key is available on the front panel.
:LIST:STIMulus:MDATa?
Syntax :LIST:STIMulus:MDATa?
Description Returns actual stimulus values for a list sweep with two parameters. This
command allows you to know the actual applied values after the auto polarity
function works. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:LIST:STIMulus:TYPE
Syntax :LIST:STIMulus:TYPE
{FREQuency|VOLTage|CURRent|BVOLtage|BCURrent|DCSVoltage},{NONE|IRA
Nge|FREQuency|VOLTage|CURRent|BVOLtage|BCURrent|DCSVoltage}
:LIST:STIMulus:TYPE?
Description Selects the parameter types for the list sweep with two parameters. If you do
not want to specify the second parameter, enter NONE for it. When you do
specify the second parameter, its measurement range can be set.
Parameter
First Parameter
FREQuency (Preset value) Selects frequency as the first stimulus parameter
VOLTage Selects voltage as the first stimulus parameter
CURRent Selects current as the first stimulus parameter
BVOLtage Selects bias voltage as the first stimulus parameter
BCURrent Selects bias current as the first stimulus parameter
DCSVoltage Selects DC source voltage as the first stimulus
parameter
Second Parameter
NONE (Preset value) Selects no second stimulus parameter. The same operation is
performed as the :LIST:FREQuency, :LIST:CURRent and
so on according to the first stimulus parameter.
IRANge Selects impedance range as the second stimulus parameter.