Keysight E4980A/AL Precision LCR Meter 365
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key [Save/Recall] - SAVE DATA - SAVE & STOP
This key is a combination of both :MEMory:READ? and :MEMory:CLEar.
:MMEMory:DELete[:REGister]
Syntax :MMEMory:DELete[:REGister] <numeric>
Description Deletes the state from the memories. Numbers 0 to 9 are located in the
internal memory, while Nos. 10 to 19 are situated in the external USB memory.
(No Query)
Parameter
Equivalent key [Save/Recall] - No. - DELETE
:MMEMory:LOAD:STATe[:REGister]
Syntax :MMEMory:LOAD:STATe[:REGister] <numeric>
Description Loads the state from the memories. Numbers 0 to 9 are located in the internal
memory, while Nos. 10 to 19 are situated in the external USB memory. (No
Query)
Parameter
Equivalent key [Save/Recall] - No. - RECALL
:MMEMory:STORe:STATe[:REGister]
Syntax :MMEMory:STORe:STATe[:REGister] <numeric>
Description Stores the state to the memories. Numbers 0 to 9 are located in the internal
memory, while Nos. 10 to 19 are situated in the external USB memory. Number
10 is automatically recalled. (No Query)
<Numeric>
Range 0 to 19
Resolution 1
<Numeric>
Range 0 to 19
Resolution 1