Keysight E4980A/AL Precision LCR Meter 163
System Configurations
SYSTEM CONFIG Page
5-
Configuring the GPIB Address
Functional Description
Before you can control the E4980A/AL by issuing GPIB commands from an
external controller connected via its GPIB connector, you have to configure the
GPIB address of your E4980A/AL. For information on the concept and
implementation of automatic configuration by use of GPIB, see “GPIB remote
control system” on page 226.
To configure the GPIB address:
Step 1. Press [System].
Step 2. Press the SYSTEM CONFIG softkey.
Step 3. Use the cursor keys to select the GPIB ADDR field.
Step 4. Enter a value with the entry keys in the range of 0 through 30.
MINUTE INCR + Increments the minutes in steps of 1.
MINUTE DECR - Decrements the minutes in steps of 1.
SECOND INCR + Increments the seconds in steps of 1.
SECOND DECR - Decrements the seconds in steps of 1.
When you enter the hour with the entry keys, use a value within the range
of 0 (midnight) through 23 (11:00 pm).
Softkey Description