368 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Description Enables the DC Source. The DC source is automatically turned OFF after
recalling the state from memory. When DC Source is set to ON, DC Source
voltage that has been set with
:SOURce:DCSource:VOLTage[:LEVel] is output.
Parameter
Equivalent key [DC Source]
:SOURce:DCSource:VOLTage[:LEVel]
Syntax :SOURce:DCSource:VOLTage[:LEVel] <numeric>
:SOURce:DCSource:VOLTage[:LEVel]?
Description Sets the DC source’s voltage level
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - DC SRC
:STATus:OPERation:CONDition?
Syntax :STATus:OPERation:CONDition?
Description Returns the value of the Operation Status Condition register. (Query Only)
Equivalent key No equivalent key is available on the front panel.
This is available when option 001 is installed.
Description
ON or 1 Enables the DC Source
OFF or 0 (Preset value) Disables the DC Source
This is available when option 001 is installed.
<Numeric>
Range -10 to 10
Preset value 0
Unit V
Resolution 1 m