EasyManua.ls Logo

Keysight E4980AL - Current[:Level]; Display:cclear; Display:enable

Keysight E4980AL
529 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4980A/AL Precision LCR Meter 343
SCPI Command Reference
E4980A/AL commands
10-
[Meas Setup] - CORRECTION - SHORT A
[Meas Setup] - CORRECTION - SHORT B
[Meas Setup] - CORRECTION - LOAD A
[Meas Setup] - CORRECTION - LOAD B
:CURRent[:LEVel]
Syntax :CURRent[:LEVel] <numeric>
:CURRent[:LEVel]?
Description Sets the current level for the measurement signal. If the signal is set using
:VOLTage[:LEVel], the query command returns error No. 230.
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - LEVEL
[Meas Setup] - MEAS SETUP - LEVEL
:DISPlay:CCLear
Syntax :DISPlay:CCLear
Description Clears errors or caution messages from the display. (No Query)
Equivalent key No equivalent key is available on the front panel.
:DISPlay:ENABle
Syntax :DISPlay:ENABle {ON|OFF|1|0}
:DISPlay:ENABle?
Description Enables display updates
<Numeric>
Range
0 to 100m
1
Unit A
Resolution 1
1. Depending on the DC Bias level.

Table of Contents

Other manuals for Keysight E4980AL

Related product manuals