10-
Equivalent key [Display Format] - MEAS DISPLAY - RANGE
[Meas Setup] - MEAS SETUP - RANGE
:FUNCtion:IMPedance[:TYPE]
Syntax :FUNCtion:IMPedance[:TYPE]
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS|LPD|LPQ|LPG|LPRP|LPRD|LSD|LSQ|LSRS
|LSRD|RX|ZTD|ZTR|GB|YTD|YTR|VDID}
:FUNCtion:IMPedance[:TYPE]?
Description Selects the measurement function.
Parameter
Description
CPD(Preset value) Sets the Impedance Parameter Type to “Cp-D”
CPQ Sets the Impedance Parameter Type to “Cp-Q”
CPG Sets the Impedance Parameter Type to “Cp-G”
CPRP Sets the Impedance Parameter Type to “Cp-Rp”
CSD Sets the Impedance Parameter Type to “Cs-D”
CSQ Sets the Impedance Parameter Type to “Cs-Q”
CSRS Sets the Impedance Parameter Type to “Cs-Rs”
LPD Sets the Impedance Parameter Type to “Lp-D”
LPQ Sets the Impedance Parameter Type to “Lp-Q”
LPG Sets the Impedance Parameter Type to “Lp-G”
LPRP Sets the Impedance Parameter Type to “Lp-Rp”
LPRD
1
Sets the Impedance Parameter Type to “Lp-Rdc”
LSD Sets the Impedance Parameter Type to “Ls-D
LSQ Sets the Impedance Parameter Type to 'Ls-Q'
LSRS Sets the Impedance Parameter Type to “Ls-Rs”
LSRD
1
Sets the Impedance Parameter Type to “Ls-Rdc”
RX Sets the Impedance Parameter Type to “R-X”
ZTD Sets the Impedance Parameter Type to “Z-thd”
ZTR Sets the Impedance Parameter Type to “Z-thr”