Keysight E4980A/AL Precision LCR Meter 135
Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
4-
Figure 4-14 LIMIT TABLE SETUP Page
Parameter Swap Feature
Functional Description
The parameter swap feature exchanges the primary and secondary parameters
in the FUNC field with each other. For example, when the measurement
function is Cp-D and you use the parameter swap feature, the measurement
function is changed to D-Cp (see
Figure 4-15); after the swap, the comparator
uses up to nine sets of limits for parameter D and one set of limits for
parameter Cp.