150                           Keysight E4980A/AL Precision LCR Meter 
Configuring Measurement Conditions (Display and Function Related Settings)
LIST SWEEP SETUP Page
4-
Sweep Points and Limit Modes
Functional Description
The list sweep measurement feature supports up to 201 sweep points as well 
as measurement limit values. Figure 4-22 shows the fields you can use to 
configure the sweep points, limit parameters, and upper and lower limit values. 
Figure 4-22  Sweep point fields
1 sweep points
2 limit parameters
3 lower limit values
4 upper limit values
If you configure only the lower limit value for a sweep point, DUTs that are 
below the lower limit value for the selected primary or secondary 
parameter are evaluated as LOW. If you configure only the upper limit 
value for a sweep point, DUTs that are above the upper limit value for the 
selected primary or secondary parameter are evaluated as HIGH. 
If you specify the limit parameter for a sweep point but omit both lower 
and upper limit values, DUTs always pass the sweep point evaluated as IN.