340 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Parameter
Equivalent key [Meas Setup] - CORRECTION - MODE - MULTI
[Meas Setup] - CORRECTION - REF A
[Meas Setup] - CORRECTION - REF B
:CORRection:SPOT[1-201]:OPEN[:EXECute]
Syntax :CORRection:SPOT[1-201]:OPEN[:EXECute]
Description Executes OPEN correction at the specified measurement point. (No Query)
Equivalent key [Meas Setup] - CORRECTION - FREQ - MEAS OPEN
:CORRection:SPOT[1-201]:SHORt[:EXECute]
Syntax :CORRection:SPOT[1-201]:SHORt[:EXECute]
Description Executes SHORT correction at the specified measurement point. (No Query)
Equivalent key [Meas Setup] - CORRECTION - FREQ - MEAS SHORT
:CORRection:SPOT[1-201]:STATe
Syntax :CORRection:SPOT[1-201]:STATe {ON|OFF|1|0}
:CORRection:SPOT[1-201]:STATe?
Description Enables the specified measurement point.
Parameter
Equivalent key [Meas Setup] - CORRECTION - FREQ - ON
[Meas Setup] - CORRECTION - FREQ - OFF
<reference value for primary>, <reference value for
secondary>
Preset value 0
Unit Depends on :CORRection:LOAD:TYPE
Description
ON or 1 Enables the specified measurement point
OFF or 0 (Preset value) Disables the specified measurement point