Keysight E4980A/AL Precision LCR Meter 429
Troubleshooting
Check Items When Trouble Occurs
13-
Perform resistance measurement of 100 to 1 k DUT by using the
E4980A/AL’s fixture after turning off OPEN/SHORT/LOAD on the
CORRECTION page and setting cable length to zero. If a measured value
close to the DUT’s value is displayed, there is the possibility of correction
failure or inappropriate cable length.
For those cases, perform the correction with proper OPEN/SHORT/LOAD or
with proper fixture/cable.
o When the MULTI compensation is ON, check that the channel and the
definition method of the LOAD standard value are selected correctly.
Saving to USB memory fails
Some USB memory does not work on the E4980A/AL with firmware A.xx.xx.
See the section on “USB interface port” in data sheet.
An error message or warning message is displayed
For information on error messages and warning messages, refer to “Error
Messages” on page 435.