122 Keysight E4980A/AL Precision LCR Meter
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
4-
You can specify up to 201 measurement points.
Step 5. Using the cursor keys, select the FREQ field.
Step 6. Enter the frequency using the entry keys. When you have entered the
frequency, the softkey labels change to unit labels (Hz, kHz, MHz).
Step 7. Press the ON softkey.
Step 8. Connect the UNKNOWN terminal and the text fixture with no DUT connected.
Step 9. Press the MEAS OPEN softkey.
During the measurement, an “OPEN measurement in progress” message is
shown on the display.
When the measurement has finished, the OPEN measurement in progress”
message disappears.
Softkey Description
INCR ++ Increments the measurement point number in steps of 10.
INCR + Increments the measurement point number in steps of 1.
DECR - Decrements the measurement point number in steps of 1.
DECR -- Decrements the measurement point number in steps of 10.