124 Keysight E4980A/AL Precision LCR Meter
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
4-
To specify frequency points and carry out short correction:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the SPOT No. field.
Step 4. Enter the measurement point number using the softkeys or entry keys. If you
have used the entry keys to enter the value, the softkey labels change to unit
labels (x1).
You can specify up to 201 measurement points.
Step 5. Using the cursor keys, select the FREQ field.
Step 6. Enter the frequency using the entry keys. When you enter the frequency, the
softkey labels change to unit labels (Hz, kHz, MHz).
Step 7. Press the ON softkey.
Step 8. Connect the UNKNOWN terminal and the test fixture and short-circuit the
HIGH and LOW test terminals.
Step 9. Press the MEAS SHORT softkey.
• During the measurement, a “SHORT measurement in progress” message is
shown on the display.
• When the measurement has finished, the “SHORT measurement in
progress” message disappears.
Softkey Description
INCR ++ Increments the measurement point number in steps of 10.
INCR + Increments the measurement point number in steps of 1.
DECR - Decrements the measurement point number in steps of 1.
DECR -- Decrements the measurement point number in steps of 10.