Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
4-
• Enable or disable the open/short/load correction data at each frequency
point specified in the FREQ field.
The following tables show how correction data elements are related to each
other when the correction features are enabled in the OPEN, SHORT, and LOAD
fields. Correction data elements are selected based on the number of
measurement points.
• When test frequency point is equal to the frequency point specified in FREQ
field:
• When test frequency point is NOT equal to the frequency point specified in
FREQ field:
Correction Frequency point specified in FREQ field
ON OFF
OPEN: ON
SHORT: ON
LOAD: ON
User-specified
User-specified
User-specified
Interpolated
Interpolated
X
Correction Frequency point specified in FREQ field
ON OFF
OPEN: ON
SHORT: ON
LOAD: ON
Interpolated
Interpolated
X
Interpolated
Interpolated
X
User-specifie
d
Correction data at your specified frequency point are used.
Interpolated Correction data generated by interpolation are used.
X Correction is not performed even when correction features are
enabled in OPEN, SHORT, and LOAD fields.
When you specify a frequency point whose frequency value is identical to
that of an existing frequency point and turn on the correction feature for
both points, the frequency point with the smaller measurement point
number (SPOT No. field) is selected as the correction data.