180 Keysight E4980A/AL Precision LCR Meter
Save/Recall
Saving/Recalling Instrument Configuration States
6-
Step 5. Press the SAVE softkey to save configuration states into the USB memory.
To recall configuration states from USB memory:
Step 1. Press [Save/Recall].
Step 2. Use the cursor keys to select the MEDIA field.
Step 3. Press the EXT softkey.
Step 4. Use the cursor keys to select the No. field (10 through 19) for your desired
register.
Step 5. Press the RECALL softkey to recall configurations from the USB memory.
If you attempt to recall configuration states under any of the following
conditions, an “Incompatible state file” warning message is displayed.
• The configuration states were saved under a different firmware version.
• The configuration states were saved from another E4980A/AL with a
different option(s) equipped.
If you attempt to recall configuration states under the following condition, a
“No data to load” error message is displayed.
• The checksum has changed.
If you have selected the No. field for a register that already contains
configuration states, the new configuration states overwrite the existing
ones.
State files are automatically assigned file names 10.sta through 19.sta, and
you cannot change the file names.
DC bias and DC source settings are automatically turned OFF.