266 Keysight E4980A/AL Precision LCR Meter
Overview of Remote Control
Data Transfer
8-
The ASCII data output format on the LIST SWEEP page is shown below. The
data loop is repeated according to the number of sweep points.
Figure 8-22 ASCII Format 2 (List Sweep)
The <DATA A>, <DATA B>, and <STATUS> formats are the same as the formats
on the MEAS DISPLAY, BIN No., or BIN COUNT page. Therefore, only the
<IN/OUT> format will be described here.
<IN/OUT> format:
The <IN/OUT> data shows the result of the list sweep’s comparator
function.
-1: LOW
0: IN
+1: HIGH
When the comparator function of the list sweep measurement is not used,
the <IN/OUT> data output result is 0 (zero).
The data output formats for <IN/OUT> use the two-ASCII character
fixed-length format as follows.
SN (S:+/-, N: 0 to 1)