Millipore System Documentation Set
116 Operating and Maintenance Manual Publication XITXSP121 P Rev.A
Entry Value
required in
field?
Yes/No
Mandatory or
optional to modify
field when
programming a new
test?
Instructions Notes
Filter Catalogue
Number
Yes Mandatory Location:
Tab F2
Select a filter
from the list.
Locate filters by either
scrolling through the list
using the arrow bars or
by using the “Quick
Search” feature. The
Quick Search feature
allows the user to type
in the first few
characters(up to four) of
the filter catalogue
number and will go
directly to the filter(s)
that correspond to
those characters.
Test Type
Yes Mandatory Location:
Tab F2
Select a test
from the list.
When a filter is chosen
the allowable tests for
that filter appear in the
Test Type dialog box.
Number of
Cartridges
Yes Optional
Default value is
commonly used.
Location:
Tab F2
Type in the
number of filters
that will be
tested with this
test program.
A default value of 1 will
appear. This only needs
to be modified if more
filters are being tested.
Housing Type
No Optional Location:
Tab F3
Type in any text
description that
is desired.
This is a text field that is
for descriptive
purposes.
Preset Upstream
Gas Volume
No Optional Location:
Tab F3
Click on the
check box and
then enter the
upstream
volume in the
dialog box.
This is used to bypass
the upstream gas
volume sizing, which
will not be performed if
this selection is chosen.
Instead the program will
use the Upstream Gas
Volume that is entered.