Integritest Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual Test Editor 117
Entry Value
required in
field?
Yes/No
Mandatory or
optional to modify
field when
programming a new
test?
Instructions Notes
Test Time
Yes Optional
Default value is
commonly used.
Location:
Tab F4
Type in a new
default test time
if desired.
Do not enter a test time
of less than 6 minutes.
This may cause the test
to terminate before a
result is calculated by
terminating prior to
completion of the
upstream volume sizing.