13-4
13.2 Procedures for machine operation
1. Turn on power breaker of the machine.
2. Execute prerinsing operation.
3. Execute dialysis.
4. Execute postrinsing.
5. Turn off power breaker of the machine.
13.2.1 Machine SW ON
(1) Check correct connection of power plug to receptacle.
(2) Check no lighting of power lamp.
(3) Turn on power breaker placed at rear side of the machine.
z Power lamp is ON.
z Information is displayed on operation panel.
SINGLE PATIENT DIALYSIS MACHINE
SURDIAL TYPE
NIPRO CORPORATION RAM:OK
ROM Ver.3.00/04 ROM:OK
F・1
F・2
F・3
F・4
F・5
NOTE
z When recovery from power failure is notified, press [MUTE] SW and [RESET] SW to reset
the machine functions.
13.2.2 Prerinse
(1) Check ON of [DIALYSIS] SW lamp.
(2) Press [DIALYSIS] SW.
z [DIALYSIS] SW lamp is OFF.
z Rinse menu screen is displayed.
[RINSE MENU]
RINSE1:WATER RINSE3:AUTO1 C&D
RINSE2:CHEMICAL RINSE4:AUTO2 C&D
RINSE1 RINSE2 RINSE3 RINSE4 DISP.
F・1
F・2
F・3
F・4
F・5
(3) Follow the procedures specified in Chapter 14 to rinse the machine.
CAUTION
z Rinse the machine before starting dialysis to prevent cross infection among patients.