21-10
21.4. Specifications for options
21.4.1. Arterial pressure sensor
Setting arterial pressure sensor unit to the machine can monitor arterial pressure.
Specification Required condition
Allowable set range
-400+400mmHg
No
Allowable indication
range
Digital indication -400 to
+400mmHg
Indicator indication - -400
+400mmHg
No
Accuracy of indicated
Value
Actual pressure±10mmHg
or ±10 (F.S.) whichever greater
Excluding indicator
21.4.2. Backup unit of blood pump for power failure
Blood pump operation is enabled during power failure by setting backup unit of blood pump
for power failure to the machine.
21.4.2.1. Machine specification
Specification Condition
Blood pump flow
rate during
power failure
0 to 185mL/min Use of blood line:
D=φ8.00×φ12.00±0.15mm
L= 210mm
Full charge time
48 hours
Time which is taken full
charged continuous power
on.
Continuous
operation time
About 90 minutes at a flow rate of
185mL/ min
When new battery is full
Charged
The specification may not be satisfied according to battery and operation conditions.
21.4.2.2. Energy cell specification
Specification Condition
Type
Ni-Cd battery No
Voltage
12V (1.2V×10)
No
Capacity
1100mAh No
Ambient
+15 to +35
No