13-5
13.2.3 Dialysis
(1) Check OFF of [DIALYSIS] SW lamp.
(2) Press [DIALYSIS] SW.
z [DIALYSIS] SW lamp is OFF.
z Dialysis screen is displayed.
[DIALYSIS]
VP: 0mmHg DP: 5mmHg COND: 0.0mS/cm
AP: 0mmHg TMP: -5mmHg TEMP:20.0゜C
S.N. SQHD PRIM TEST DISP.
(3) Follow the procedures specified in Chapter 15 or 16 to execute dialysis.
13.2.4 Postrinse
(1) Check ON of [DIALYSIS] SW lamp.
(2) Press [DIALYSIS] SW.
z [DIALYSIS] SW lamp is OFF.
z Rinse menu screen is displayed.
[RINSE MENU]
RINSE1:WATER RINSE3:AUTO1 C&D
RINSE2:CHEMICAL RINSE4:AUTO2 C&D
RINSE1 RINSE2 RINSE3 RINSE4 DISP.
(3) Follow the procedures specified in Chapter 14 to rinse the machine.
CAUTION
z Rinse the machine after dialysis termination to prevent cross infection among patients,
z Rinse the machine after disconnection of patient from it.
13.2.5 Machine power breaker OFF
(1) When postrinsing is completed, press [DIALYSIS] SW to display dialysis screen.
(2) Turn off power breaker placed at rear side of the machine.
z Power lamp is OFF.
z Information on operation panel is OFF.
NOTE
z The machine can be left under power breaker OFF until the next dialysis. It is
recommended, however, to turn off the breaker when the machine is not operated for long
time.