16-7
16.3.4.2. Setting of syringe operation stop time
By the setting, automatic syringe operation stop is enabled at a set time before UF
completion and the stop is indicated by error message "E61 COMPLETION OF SYRINGE.P
INFUSION".
(1) Press [F・5] SW to display settingⅠscreen.
(2) Press [F・3] or [F・4] SW to blink "3.S.STP(min)".
(3) Press [F・1] or [F・2] SW to change set value.
(4) Check the changed set value.
[SETTING1]
1.TEMP (°C):37.0 2.BAR GRAPH :DP
3.S.STP(min): 0 4.BOLUS (ml):0.0
+ - < > DISP.
16.3.4.3. Setting of fast-forward volume
By pressing [F.F.] SW during syringe pump operation, pusher moves at fast forward speed.
Pusher moves a specified speed to perform bolus infusion followed by automatic returning
to the
normal infusion after infusing a set infusate volume.
(1) Press [F・5] SW to display settingⅠscreen.
(2) Press [F・3] or [F・4] SW to blink "4.BOLUS(ml)".
(3) Press [F・1] or [F・2] SW to change set value.
(4) Check the changed set value.
[SETTING1]
1.TEMP (°C):37.0 2.BAR GRAPH :DP
3.S.STP(min): 0 4.BOLUS (ml):0.0
+ - < > DISP.
z0.0 :Bolus infusion during pressing [F.F] SW only until reaching to maximum
infusion
as 5 mL. Return regular infusion automatically when bolus infusion is
reached to
maximum infusion.
z0.5-5.0 :Bolus infusion of set infusate volume
CAUTION
z Syringe pump may not stop infusing after releasing [F.F] SW by some malfunction.
In this case, stop syringe pump by turn OFF the breaker of the machine and then
interrupt
treatment. Machine must be disconnected from patient.