21.2. Performance of machine
21.2.1. Dialysate flow rate
Specifications Condition
Dialysate flow
Rate
Min 300mL/min
Rating500mL/min
Max 600mL/min
UFR is 2.70L/h when dialysate
flow rate is 300mL/min.
21.2.2. Deaeration
Specifications Condition
Deaeration
System
Negative pressure deaeration system No
Gas dissolved in
dialysate
Max130mmHg
When PO
2
is 200mmHg after
deaeration at a water temp.
of 37.
21.2.3. UFR control
Specifications Condition
UFR control
System
Closed capacity difference control by
VCS ( Viscous Control System)
No
Possible setting
Range
0 0.105.00L/h UFR is 2.70L/h when
dialysate flow rate is
300mL/min.
Accuracy of
UFR
±30g/h at a UFR of 0.10-5.00L/h
If machine settings at release
from factory are maintained
21.2.4. Temperature control
Specifications Condition
Possible setting
Range
30.040.0
0.1 steps
Supply water temp. should
be lower than set dialysate
temp. by 5.
Ambient temp. should be
lower than set dialysate
temp. by 5.
Accuracy of
Indicated temp.
Indicated dialysate temp.:
Within mean temp.±0.8
Mean temp.: Mean of
dialysate temp. for one
minute