13-1
Chapter 13
OPERATION OF MACHINE
13.1 Check and action before turning on power breaker
Before starting daily operation of the machine, check the following items specified in the
below Table and correct non-conformance immediately if detected.
WARNING
z If non-conformance can not be corrected, DO NOT operate the machine.
No. Item to be checked
Paint out
1 Installation conditions
Ambient temperature should be 1535.
Relative humidity should be 3580.
Atmospheric pressure should be 7951062hPa.
Spaces should be 5cm above top of the machine,
10cm from each of right and left sides of it and 20cm
from rear side of it.
Ventilation port of ≧φ20cm should be provided or at
least one side other than the machine bottom should be
exposed to atmosphere.
OKNG
OKNG
OKNG
OKNG
OKNG

Table of Contents