21-5
21.2.5. Conductivity control (BICARBONATE
Specifications Condition
Dialysate
preparation
System
Fixed volume mixing by VCS (Viscous
Control System)
No
Possible input range
of dialysate
concentrate dilution
ratio based on
concentrate supplier
data
A concentrate : 1/30.00 to 1/44.00
(at 0.01 steps of denominator)
B concentrate : 1/14.00 to 72.00
(at 0.01 steps of denominator)
No
Possible input range
of concentration
base on concentrate
supplier data
A concentrate : 70 to 130 mmol/L
(at 1 mmol/L steps)
B concentrate : 10 to 90 mmol/L
(at 1 mmol/L steps)
No
Possible setting
range of conductivity
A+W : 70 to 130mmol/L
(at 1mmol/L steps)
TOTAL : 130 to 160mmol/L
(at 1mmol/L steps)
No
21.2.6. Conductivity control (ACETATE)
Specifications Condition
Dialysate
preparation
System
Fixed volume mixing by VCS (Viscous
Control System)
No
Possible input range
of dialysate
concentrate dilution
ratio based on
concentrate supplier
data
1/30.00 to 1/44.00
(at 0.01 steps of denominator)
No
Possible input range
of concentration
base on concentrate
supplier data
90 to 160mmol/L
(at 1mmol/L steps)
No
Possible setting
range of conductivity
90 to 160mmol/L
(at 1mmol/L steps)
No