11.3.3 Method to set dilution rate of disinfectant concentrate
It is necessary to release protect for setting based on the procedure specified in “Service
Manual” to set dilution rate of disinfectant concentrate.
(1) Press [F5] SW to display maintenance mode menu screen.
(2) Press [F3] or [F4] SW to display "MODE7" and press [F1] SW.
[MAINTE MENU(4/5)]
MODE7:SYSTEM
MODE8:SERVICE FREQUENCY
MODE7 MODE8 P.UP P.DOWN DISP.
F・1 F・2 F・3 F・4 F・5
(3) Press [F3] or [F4] SW to blink "4.DISINFECTION RATIO".
(4) Press [F1] or [F2] SW to set dilution rate of disinfectant concentrate.
[MAINTE 7]
4.DISINFECTION RATIO :1/300
5.ACID RATIO 1/30
DOWN MENU
F・1 F・2 F・3 F・4 F・5
(5) Press [F5] SW to return to maintenance menu screen.
11.3.4 Check of residual disinfectant
WARNING
z If disinfectant remains in the machine even though the amount is small, patient life may
be threatened.
z Execute water rinse for 30 minute or more after disinfection. The time, however, is
changed according to diluted disinfectant concentration and dialysate flow rate. It is
recommended to check residual disinfectant amount using test paper or reagent.