10-4
10.2.2 Table of alarms
Alarm
No.
Alarm name
Detection
condition
Detection
process
Audible alarm
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
E01 RECOVER FROM
EMERGENCY STOP
Return to normal after
power failure caused
during sequence
execution
Rinse
Preparation
Preparation end
Dialysis
Recovery
(3 times)
Execution of
sequence before
error detection
Automatic return to normal
Becoming error free of set upper and lower
VP/ AP/ DP alarm limits by alarm actuation
E02 T1 TEMPERATURE
SENSOR ERROR
T1 temperature exceeding
measurable temperature
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(7 times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
Continuance
Same as above
Same as above
Same as above
Same as above
Fluid line stop
Same as above
Same as above
Same as above
Return to normal by pressing [RESET] SW
E03 T2 TEMPERATURE
SENSOR ERROR
T2 temperature exceeding
measurable temperature
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(7 times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
Continuance
Same as above
Same as above
Same as above
Same as above
Fluid line stop
Same as above
Same as above
Same as above
Return to normal by pressing [RESET] SW
During rinse process, upper value of
measurable range could not be detected.
E04 T3 TEMPERATURE
SENSOR ERROR
T3 temperature exceeding
measurable temperature
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(7 times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
Continuance
Same as above
Same as above
Same as above
Same as above
Fluid line stop
Same as above
Same as above
Same as above
[Return to normal by pressing [RESET] SW
E05 DIALYSATE
TEMPERATURE
MAX ALARM
Dialysate temperature (T3)
of set upper alarm limit
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
Auto bypass
Heater OFF
Return to normal by pressing [RESET] SW
Auto reset after release from detection
condition for 10 sec