15.11. Completion of dialysis (recovery process)
1. Check UF completion.
2. Start blood recovery.
3. Execute blood recovery.
4. Remove coupler.
5. Clear away.
6. Complete blood recovery.
15.11.1. Check of UF completion
(1) Check that “PREFORMED TARGET UF " is alerted.
(2) Check conformance of "UF GOAL" value to "UF REMOVED" value.
(3) When "MINIMUM UF" is set, check that “MINIMUM UF” function is executed.
(4) Check that fluid lines are functioning normally.
(5) Press [MUTE] SW and [RESET] SW to mute buzzing to alert “
PREFORMED TARGET UF " and
erase the alarm message respectively.
15.11.2. Start of blood recovery
(1) Press [PREP.] SW once.
z[PREP.] SW lamp is ON.
z[UF DIALYSE] SW lamp is OFF.
zFluid lines are functioning normally.
zUF is stopped.