21-7
21.2.10. Blood pump
Specifications Condition
Blood pump
type
Three roller type
Manual gap adjustment
(range: 2.5 to 3.5 mm from housing
inside)
No
Flow rate range When large sized tube is used :
30 to 500 mL/min (factory
set)
When small sized tube is used :
30 to 300 mL/min
Large sized tube:
φ8.00 × φ12.00 ± 0.15 mm
Length : ≥ 210 mm
Small sized tube:
φ 6.35 × φ 9.75 ± 0.15 mm
Length : ≥ 210 mm
Accuracy of
flow rate
Set flow rate ± 10 %
Min. inlet pressure :
-200 mmHg
Max. outlet pressure :
+500 mmHg
21.2.11. Syringe pump
Specification Required condition
Set infusion rate 0.1 to 9.9mL/h
No
Accuracy of
Infusion rate
Mechanical error ±1%
Infusion rate error ±10%
Nipro 20mL syringe
Overload
detection
Detection of overload: 54 to 60N
When “2 Nipro 20mL
syringes” is set
Bolus infusion
rate
About 1800mL/h
Nipro 20mL syringe
Bolus infusion
Volume
0.5 to 5.0mL
0.0mL: disabling of bolus infusion
Nipro 20mL syringe
Syringe pump
Operation stop
0 to 60min
0min: disabling of the operation stop
No
Applicable
syringes
Ones of cross-section ratio between 50
and 200%
Nipro 20mL syringe: 100%
No