14-3
14.3 Connection of concentrates
CAUTION
z Connect each concentrate connector and rinse port securely. If not so, troubles such as
water leak and air aspiration may be caused.
z Concentrate type(s) and concentrate connector type(s) are defined by use or no use of
acid rinse port unit (option) and rinse mode to be executed. Take care of this when
connection is executed.
z Use disinfectant and acetic acid concentrates of specified concentrations.
14.3.1 Connection of concentrates when acid rinse port (option) is not used
WARNING
z When acid rinse port (option) is not used, aspiration port of disinfectant or acetic acid
concentrate can be used as connector of disinfectant concentrate container. Check
conformance of concentrate type to rinse mode to be executed.
14.3.1.1 Connection at water rinse
(1) Connection of any concentrate is not required. Connect all concentrate connectors to
respective rinse ports or engage them with connector receivers.
(2) Check correct connector connection.
14.3.1.2 Connection at disinfection/rinse
(1) Connect disinfectant concentrate connector placed at rear side of the machine to
disinfectant concentrate container.
(2) Connect other connectors to respective rinse ports or engage them with connector
receivers.
(3) Check correct connector connection.
14.3.1.3 Connection at acid rinse
(1) Connect disinfectant concentrate connector placed at rear side of the machine to acetic
acid concentrate container.
(2) Connect other connectors to respective rinse ports or engage them with connector
receivers.
(3) Check correct connector connection.