15-13
15.7. Check of dialysate conductivity
(1) Prepare instrument allowing accurate measurement of conductivity within range from
10.0 to 18.0mS/cm.
(2) Prepare syringe used to take dialysate sample.
(3) Take the sample from sampling port of the machine using syringe.
(4) Measure conductivity of the sample.
(5) Check whether conductivity conforms to requirement.
15.8. Setting of UF related parameters
(1) Check that indicated “UF REMOVED” is 0.00.
(2) Set target UF volume by pressing [△] or [▽] SW for setting of “UF GOAL”.
(3) Set UFR by pressing [△] or [▽] SW for setting of “UF RATE”. By the setting, “TIME
LEFT” value is computed and indicated automatically.