18.2.6.2. Zero revision of blood leak sensor
Adjust the zero revision of blood leak detector when “6.BLD AUTO ZERO” value less
than
-150ppm in maintenance mode 1 screen during preparation process.
CAUTION
z Perform water rinse or disinfection before you calibrate it.
It caused of display error if you calibrate it even blood leak detector is dirty.
(1) Press [Prep.] SW to shift to preparation process.
(2) Check preparation time pass over 5 minutes.
(3) Turn adjustment volume 1 counterclockwise.
Adjust “6.BLD AUTO ZERO” value to -40~-30ppm.
Adjustment value Adjustment point
Maintenace mode 1 : BLD AUTO
ZERO
-40~-30ppm
Adjustment volume
1
(※1) Check no bubble inside cartridge of blood leak detector.
(※2) DO NOT turn adjustment volume 1 counterclockwise too much.
BLD value is displayed lower than actual value.
18-10
[MAINTE 1]
6.BLD AUTO ZERO : ―30
7.BLD SENS VALUE : ――――――
UP DOWN MENU
F・1 F・2 F・3 F・4 F・5
Fig.18-9 Maintence mode 1 screen
BLD cover
Blood leak detector