15-3
15.1.2.2. Acetate dialysate
(1) Press [F・5] SW to display settingⅡscreen.
(2) Press [F・3] or [F・4] SW to blink "3.AC(W+AC)".
(3) Press [F・1] or [F・2] SW to change setting.
(4) Check correct setting.
[SETTING2]
1.BC(W+A+B) :140 2.BC(W+A) :110
3.AC(W+AC) :140 4.MODE :BC1
+ - < > DISP.
z3.AC(W+AC) :Setting of final concentration of acetate dialysate by mmol/l unit
15.1.3. Setting of dialysate temperature
(1) Press [F・5] SW to display settingⅠscreen.
(2) Press [F・3] or [F・4] SW to blink "1.TEMP(℃)".
(3) Press [F・1] or [F・2] SW to change setting.
(4) Check correct setting.
[SETTING1]
1.TEMP (°C):37.0 2.BAR GRAPH :DP
3.S.STP(min): 0 4.BOLUS (ml):0.0
+ - < > DISP.