16-2
16.1. Setting of dialysate related parameters
1. Set dialysis mode.
2. Set dialysate concentration.
3. Set dialysate temperature.
16.1.1. Setting of dialysis mode
(1) Press [F・5] SW to display settingⅡscreen.
(2) Press [F・3] or [F・4] SW to blink "4.MODE".
(3) Press [F・1] or [F・2] SW to change setting.
(4) Check correct setting.
[SETTING2]
1.BC(W+A+B) :140 2.BC(W+A) :110
3.AC(W+AC) :140 4.MODE :BC1
+ - < > DISP.
zBC1(BICARBONATE1) :Execution of bicarbonate dialysis
zBC2(BICARBONATE2) :Execution of bicarbonate dialysis
zAC(ACETATE) :Execution of acetate dialysis
WARNING
z Information concerning concentrate for each of BC1, BC2 and AC is found in
maintenance mode 5 screen. Set dialysis mode conforming to concentrate(s) intended to
be used.
z Both BC1 and BC2 are bicarbonate dialysis. Information not related to concentrate(s)
intended to be used may be shown in maintenance mode 5 screen. Take care of this.
16.1.2. Setting of dialysate concentration
16.1.2.1. Bicarbonate dialysate
(1) Press [F・5] SW to display settingⅡscreen.
(2) Press [F・3] or [F・4] SW to blink "1.BC(W+A+B)".
(3) Press [F・1] or [F・2] SW to change setting.
(4) Press [F・3] or [F・4] SW to blink "2.BC(W+A)".
(5) Press [F・1] or [F・2] SW to change setting.
(6) Check correct setting.
[SETTING2]
1.BC(W+A+B) :140 2.BC(W+A) :110
3.AC(W+AC) :140 4.MODE :BC1
+ - < > DISP.
z1.BC(W+A+B) :Setting of final concentration of bicarbonate dialysate by mmol/l unit
z2.BC(W+A) :Setting of concentration of [water + A concentrate] in bicarbonate
dialysate by mmol/l unit