21.3.4. Blood leak detector
Specifications Condition
Sensor type Optical detection type No
Detection Detection setting range :
50 to 500 ppm (at 10ppm steps)
Detection point 300ppm : (Ht 20 %)
No
Alerting time 5 seconds after detection No
21.3.5. Temperature monitor
Specifications Condition
Sensor type Platinum resistance thermometer No
Sensor accuracy
Actual temperature ± 0.5°C
Sensor only
Fluid temp.: ≦40℃
Alarm limit
setting
Upper alarm limit :
29.0 to 41.0 °C (at 0.1 °C steps)
Upper alarm limit accuracy :
Mean temp. ± 0.8 °C
Lower alarm limit :
29.0 to 41.0 °C (at 0.1 °C steps)
Lower alarm limit accuracy :
Mean temp. ± 0.8 °C
Mean of dialysate temp. for
one minute
21.3.6. Conductivity monitor
Specifications Condition
Sensor type Electrode type No
Reproducibility
± 1% (F.S.) At 25°C
Alarm limit
setting
Automatic alarm limit setting:
Set conductivity ± 10% (max)
Manual alarm limit setting:
1 to 10% (at 1% steps)
No difference between set level and
actually detected level
No