6-9
6.4.2. Connection of each concentrate
6.4.2.1. Connection to execute dialysis operation (before blood line connection)
(1) Load container of A concentrate (or acetate concentrate) and that of B concentrate on
respective container supports.
(2) Connect connectors of the machine for A and B concentrates to respective connectors of the
containers.
(3) Check whether connector of the machine for disinfectant concentrate is connected to rinse
port correctly.
(4) If acid rinse option is used, check whether connector for acetic acid concentrate is
connected to rinse port correctly.
CAUTION
Connect each concentrate connector, rinse port and the like securely. Otherwise, trouble
such as water leak or air aspiration may be caused.
During acetate dialysis (B concentrate is not used), connect connector for B concentrate
to corresponding rinse port placed at upper part of the machine.
6.4.2.2. Connection for water rinsing
(1) Check whether all connectors are connected to rinse ports.
CAUTION
Connect each concentrate connector, rinse port and the like securely. Otherwise, trouble
such as water leak or air aspiration may be caused.
Connect all 3 connectors of the machine for A, B and disinfectant concentrates to
respective rinse ports. If acid rinse option is used, connect acetic acid concentrate
connector to corresponding rinse port.
6.4.2.3. Connection for disinfection
(1) Connect connector placed at rear side of the machine for disinfectant concentrate to
container for disinfectant concentrate.
(2) Check whether connectors of the machine for A and B concentrates are connected to
respective rinse ports securely.
(3) If acid rinse option is used, check whether connector for acetic acid concentrate is
connected to rinse port correctly.
CAUTION
Connect each concentrate connector, rinse port and the like securely. Otherwise, trouble
such as water leak or air aspiration may be caused.