12-3
12.2.3 Rinse end process
When “rinse” process is completed, “rinse end” process is followed and fluid line function is
disabled. To repeat rinsing, turn off [DIALYSIS] SW to allow start of “rinse wait” process.
12.2.4 Dialysis wait process
When [DIALYSIS] SW is turned on, “dialysis wait” process is followed and fluid line function
is disabled.
12.2.5 Preparation process
In “dialysis wait” process, when [PREP.] SW is turned on, “preparation” process is followed
and dialysate preparation is started. During the preparation, <Standby> is displayed in screen.
The process is designed to execute fluid replacement in fluid lines.
12.2.6 Preparation end process
In “preparation” process, when the following conditionsto are satisfied, “preparation
end” process is followed. After shifting to the process, it is possible to shift to “dialysis”
process by pressing [UF DIALYSE] SW.
Dialysate conductivity within allowable alarm limits
Dialysate temperature within allowable alarm limits
Execution of 4 rounds of chamber change after start of dialysate preparation
Even in “preparation end” process, if any of the above conditions is not satisfied, dialysate
flow is auto-bypassed or any alarm is actuated, shift to “dialysis” process is disabled unless
the condition is satisfied; or the bypassing or the alarm is corrected.
12.2.7 Dialysis process
In “preparation end” or “recovery” process, when [UF DIALYSE] SW is turned on, “dialysis”
process is followed. In the process, UF is executed.
12.2.8 Recovery process
In “dialysis” process, when [PREP.] SW is turned on, “recovery” process is followed. In the
process, blood is recovered.
12.2.9 Air purge process
In “rinse wait” process, when maintenance mode 9 screen is displayed and [F1] SW is
pressed, “air purge” process is followed. In the process, air is removed from fluid lines. The
process is executed at the machine installation and the like.