16-20
16.13. Discontinuance of dialysis
By executing blood recovery, UF operation is discontinued and by shift to bypass mode
during recovery process, dialysis is discontinued. .
Discontinue UF.
Discontinue dialysis.
16.13.1. Discontinuance of UF
(1) Press [PREP.] SW once.
z[PREP.] SW lamp is ON.
z[UF DIALYSE] SW lamp is OFF.
zShift to recovery process is followed.
zFluid lines are functioning normally.
zUF is stopped.
CAUTION
z UF can be stopped by setting UFR to 0.00L/h.
z When [PREP.] SW is pressed to shift to recovery process and to disable UF, setting of all
pressure alarm limit is released. Set the limits manually if required.
z When [PREP.] SW is pressed to shift to recovery process and to disable UF , “MINIMUM
UF” function is disabled.
16.13.2. Discontinuance of dialysis
(1) Press [BYPASS] SW once.
z[BYPASS] SW lamp is ON.
zDialysate supply to dialyzer is stopped.
CAUTION
z When bypass mode is selected during dialysis, dialysate supply to dialyzer is stopped,
while UF continues.
z When bypass mode is selected during blood recovery process, dialysate supply to
dialyzer is stopped in addition to UF stop already executed.