18.2.6. Check of accuracy of blood leak sensor
18.2.6.1. Visual check of blood leak sensor
(1) Remove cartridge from blood leak detector.
(2) Confirm the dirt from surface of glass parts.
Wipe glass parts with soft cloth if it seems dirty.
(3) Confirm the dirt from inside of cartridge.
Perform water or disinfection if it seems dirty.
(4) Return cartridge to blood leak detector. Confirm the direction of O-ring when
returning cartridge.
Cartridge can be
removed by loosing
NoteIt is not necessary to remove
silicone tube.
Fig.18-8 Remove cartridge
18-9