6-6
6.3. Procedure for assembling
6.3.1. Cautions
Take care of the followings at assembling.
(a) Place the machine on horizontal and stable floor.
(b) Sufficient space to assemble should be provided.
(c) Use accessories packaged with the machine and/or designated option to assemble.
6.3.2. Attachment of I.V. pole, hanger and main lamp
Fig. 6-2 Attachment of I.V. pole
Attach the pole to the machine as shown in Fig. 6-2. Harness can be easily passed through the pole
by binding strand to the machine side connector followed by passing the connector bound strand
through the pole. Engage the connector with the lamp side connector securely.
Wires of the both connectors are colored as shown in the below Table:
Machine side (Letters on tube) Lamp side
Black or white (LAMP) Black or white
Red Red