10-2
(h) Blood or heparin pump stop
Blood or heparin pump is stopped.
(i) Clamping
When dialysate pressure related error or bubble in blood line is detected, the line
is clamped temporarily.
(j) Dialysate pressure release or air removal from dialysate
When dialysate pressure exceeds preset level, it is alerted and the pressure is
released automatically. When air is detected in dialysate lines during dialyzer
connection, the air is automatically removed.
10.1.2 Procedure for resetting
To return the machine functions to normal after removal of error causes, follow the
below procedures. See 10.2.2 “Table of Alarms” to find the error causes and the like.
Press [MUTE] SW once.
Audible alarm is muted.
[MUTE] SW lamp is OFF.
Remove error causes.
Press [RESET] SW once.
[RESET] SW lamp is OFF.
The machine functions are returned to normal.
10.1.3 Actions to respond to alarm (flow)
Error incidence
Detection of error and execution of machine functions responding to the error
Buzzing OFF by operator
Removal of error causes by operator
Resetting by operator