9-9
9.3.2.4 RINSE4: AUTO2 C&D
Press [F・3] or [F・4] SW to shift blink to desired position and press [F・1] or [F・2] SW to
change setting.
Rinse program and time can be changed.
[RINSE4] 20:15 < Wait >
WTR30→CHM30→DWL30→PRE30→-----→-----
PRERINSE START TOMORROW 10:00
+ - < > MENU
Fig. 9-8 Rinse wait screen when RINSE4 is selected
INDICATION DESCRIPTION
WTR※※ Execution of water rinse for ※※ min
CHM※※ Execution of disinfection/rinse for ※※ min
ACD※※ Execution of acid rinse for ※※ min
DWL※※ Execution of dwelling for ※※ min
PRE※※
Start of prerinse at time set by PRERINSE START
-----
No operation
PRERINSE START Setting of prerinse start time within period from this
time to 6 days after it
NOTE
z If acid rinse port unit (option) is not set to the machine, simultaneous setting of both
disinfection/rinse and acid rinse is disabled.
z If program adversely affecting the machine operation; for example, program to execute
disinfection/rinse and acid rinse in sequence; or that including no prerinse or no water
rinse step after rinse completion is set, rinsing operation can not be started even by
pressing RINSE SW.
z Refer to the option manual for setting hot rinse function. Set up hot water rinse (HWR)
and hot chemical disinfectant rinse require the machine already have hot rinse option.