13-2
No. Item to be checked
Paint out □
2 Fluid supply and drainage
・ All tubes and lines such as supply and drain tubes
should be free from damages such as folding and
bending.
・ Water supply pressure should be 0.049 ~ 0.74MPa (0.5 ~
7.5kgf/cm
2
)
・ Supply water flow rate should be ≧750mL/min.
・ Supply water temperature should be 17~30℃.
(If heat exchanger is used, it should be 5~30℃.)
・ Water quality should be suitable for dialysis use.
・ Concentrate line pressure should be -100mmHg (136cm
below inlet) to +100mmHg (136cm above inlet).
・ Drain tube end should not be contacted with waste water
level.
□OK □NG
□OK □NG
□OK □NG
□OK □NG
□OK □NG
□OK □NG
□OK □NG
3
Power supply
・ Rating of power supply should conform to that of the
machine.
・ Capacity of power supply should exceed that of the
machine.
□OK □NG
□OK NG
4
Ground connection and equipotentialization
・ Power supply should be provided with protective ground
terminal.
・ Protective ground conductors of power cord and power
plug should be connected correctly.
□OK □NG
□OK □NG
5 Connection of power supply
・ Power cord plug of the machine should be connected to
receptacle of power supply correctly.
□OK □NG
6 Cleaning
・ The machine enclosure should be cleaned.
□ OK □NG