the ratio is 100% with exception that the speed is defined in the case of bolus infusion.
When the set ratio is 200%, it moves at a half speed compared with the case that the ratio is
100% with the exception.
(1) Press [F5] SW to display maintenance mode 7 screen.
(2) Press [F3] or [F4] SW to indicate "20.MAINTENANCE MODE FREE".
[MAINTE 7]
20.MAINTENANCE MODE FREE OFF
DOWN MENU
F・1
F・2
F・3
F・4
F・5
(3) Press [F1] SW to set to "ON".
[MAINTE 7]
20.MAINTENANCE MODE FREE
DOWN MENU
F・1
F・2
F・3
F・4
F・5
(4) Instruction appears in [F1] or [F2] SW (/). Check that setting is enabled by
blinking of items to be set.
(5) Press [F3] or [F4] SW to blink "1.SYRINGE RATE".
(6) Press [F1] or [F2] SW to change set value.
[MAINTE 7]
1.SYRINGE RATE 100
2.BP STOP AFTER POWER FAIL OFF
DOWN MENU
F・1
F・2
F・3
F・4
F・5
(7) Press [F3] or [F4] SW to indicate "20.MAINTENANCE MODE FREE".
(8) Press [F1] SW to set to "OFF".
(9) Setting of syringe operation stop time
By the setting, automatic syringe operation stop is enabled at a set time before UF
completion and the stop is indicated by error message "E61 COMPLETION OF SYRINGE.P
INFUSION".
(1) Press [F5] SW to display settingscreen.
(2) Press [F3] or [F4] SW to blink "3.S.STP(min)".
(3) Press [F1] or [F2] SW to change set value.
(4) Check the changed set value.
[SETTING1]
1.TEMP (°C):37.0 2.BAR GRAPH :DP
3.S.STP(min): 4.BOLUS (ml):0.0
DISP.
F・1
F・2
F・3
F・4
F・5