15-6
the ratio is 100% with exception that the speed is defined in the case of bolus infusion.
When the set ratio is 200%, it moves at a half speed compared with the case that the ratio is
100% with the exception.
(1) Press [F・5] SW to display maintenance mode 7 screen.
(2) Press [F・3] or [F・4] SW to indicate "20.MAINTENANCE MODE FREE".
[MAINTE 7]
20.MAINTENANCE MODE FREE : OFF
UP DOWN MENU
(3) Press [F・1] SW to set to "ON".
[MAINTE 7]
20.MAINTENANCE MODE FREE : ON
+ - UP DOWN MENU
(4) Instruction appears in [F・1] or [F・2] SW (+/-). Check that setting is enabled by
blinking of items to be set.
(5) Press [F・3] or [F・4] SW to blink "1.SYRINGE RATE".
(6) Press [F・1] or [F・2] SW to change set value.
[MAINTE 7]
1.SYRINGE RATE : 100
2.BP STOP AFTER POWER FAIL : OFF
+ - UP DOWN MENU
(7) Press [F・3] or [F・4] SW to indicate "20.MAINTENANCE MODE FREE".
(8) Press [F・1] SW to set to "OFF".
(9) Setting of syringe operation stop time
By the setting, automatic syringe operation stop is enabled at a set time before UF
completion and the stop is indicated by error message "E61 COMPLETION OF SYRINGE.P
INFUSION".
(1) Press [F・5] SW to display settingⅠscreen.
(2) Press [F・3] or [F・4] SW to blink "3.S.STP(min)".
(3) Press [F・1] or [F・2] SW to change set value.
(4) Check the changed set value.
[SETTING1]
1.TEMP (°C):37.0 2.BAR GRAPH :DP
3.S.STP(min): 0 4.BOLUS (ml):0.0
+ - < > DISP.