Revision table of SURDIAL OPERATION MANUAL
We revised the following contents in operation manual.
Kindly refer the following contents in operation manual compare as manual from your side.
Page Contents Revision Previous
11-1 11.1.1.Cleaning and
disinfection of machine
enclosure
Delete cleaning procedure
using alcohol.
Remove blood and/or organic substances from
the machine enclosure using soft cloth wetted
with tepid water or water. DO NOT use organic
solvents such as thinner and benzene to
prevent patient damage.
Remove blood and/or organic substances from
the machine enclosure using soft cloth wetted
with alcohol or water. DO NOT use organic
solvents such as thinner and benzene to
prevent patient damage.
11-1 11.1.2.Cleaning and
disinfection of dialysate
concentrate connectors
and rinse ports
Delete cleaning procedure
using alcohol.
After long term use of dialysate concentrate
connectors and rinse ports, solutes in the
concentrates may isolated to form harmful
crystals and may cause the machine failure.
Clean them using cloth wetted with tepid water
or water regularly. Pull out the ports to this
side to allow easy cleaning.
After long term use of dialysate concentrate
connectors and rinse ports, solutes in the
concentrates may isolated to form harmful
crystals and may cause the machine failure.
Clean them using cloth wetted with alcohol or
water regularly. Pull out the ports to this side to
allow easy cleaning.
21-3 21.2.1.Dialysate flow
UFR is2.70L/h when dialysate flow rate is
300mL/min.
UFR is4.00L/h when dialysate flow rate is
30L/min.
0R1A01-0105