APPLIED OPERATIONS/Image Acquisition
OPERATION INSTRUCTIONS
2-79
Page
2-2-8 Image acquisition to prevent crosstalk between fluorescence (Sequential Scan)
An image which is suppressed crosstalk can be acquired sequentially by combination of
excitation laser and those image acquisition channel.
With this image capturing method, the image of a multiple-dyed specimen can be obtained
by sequentially acquiring image slice of each type of fluorescence.
1. Select the [Surface XY-Note] option button in the [Mode] group box in the [Acquire]
panel, then select <Sequen> in the list shown below.
Selecting <Sq Clp>, <Sq Rct>, or <Line Sq> enables the image acquisition of the
desired rectangle area or that of the rectangle area at desired angle, or a line
sequential scan..
Line Sequential Scan
Ordinary sequential scan is performed for every one frame for every group,
whereas line sequential scan is performed for every one line for every group.
Therefore, line sequential scan is of advantage in minimizing the time lag
between wavelength and acquiring a fewer cross-talk image.
2. Select the observation mode with the option buttons in the [Acquire] panel.

Table of Contents

Other manuals for Olympus Fluoview FV1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus Fluoview FV1000 and is the answer not in the manual?

Olympus Fluoview FV1000 Specifications

General IconGeneral
BrandOlympus
ModelFluoview FV1000
CategoryMicroscope
LanguageEnglish

Related product manuals