Getting Started FLUOVIEW/Online Help
OPERATION INSTRUCTIONS
1-46
Page
1-3-2-2 Parfocality Correction and Jog Sensitivity Adjustment
When the system use the BX or IX microscope, the parfocality correction and jog sensitivity
can be set per objective.
1. Open the [Microscope Control Panel] window.
For the method of displaying this window, see steps 1 and 2 in section 1-3-2-1,
“Configuring the Microscope ” in this section.
2. Click the <Focus Setting> button in the [Microscope Control Panel] window.
The [Focus Setting] dialog box as shown below opens.
3. When the [Focus Setting] dialog box is displayed, the objective mounted on the
microscope is selected.
To use the setup wizard to begin the parfocality correction from the value of the
objective with maximum magnification in descending order, click the <Next> button in
the [Pfcl Setup Wizard] group box.
[Objectives] group box
Click the button for the objective
subjected to parfocality correction so
that the button looks depressed.
<Pfcl> button
Depress to activate the
parfocality correction. To
deactivate, depress this
button once again.
[Jog Fine] group box
Select the jog sensitivity o
f
each objective from the
drop-down list.
[Pfcl Setup Wizard] group box
The Wizard for parfocality correction.
<Next> button
Click to execute the Parfocality Setup
(correction) Wizard.
<Import> button
Click to import the previously saved
parfocality correction and jog
sensitivity values for each objective.
<Exit> button
Click to close the [Focus
Setting] dialog box.
Display of the parfocality
correction value of each
objective.