Getting Started FLUOVIEW/Outline of LSM Observation Procedures
OPERATION INSTRUCTIONS
1-16
Page
1-2-3 Setting the LSM Light Path
1-2-3-1 Combination with Upright Microscope (BX)
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Settings]
sub-panel.
Fig. 1-5 [Settings] Sub-panel
2. Select the <LSM> button in the [Light Path] group box.
The <LSM> button looks pushed in to indicate that it is selected.
(When scanning is started while the <BI> button is selected, the LSM light path is
selected automatically. It is switched back to the visual observation automatically when
scanning completes.)
3. Push the hand switch button to (8) set
1
to be displayed in the cube display
window (1) on the reflected light fluorescence vertical illuminator