APPLIED OPERATIONS/ Protocol processor
OPERATION INSTRUCTIONS
2-152
Page
: <Get FV Status> button: Acquires each setting value of FLUOVIEW software
parameters and registers to PAPP window. The track which the parameters to
be registered has to be highlighted.
: <Set to FV> button: Sets PAPP parameters to FLUOVIEW software parameters..
The track which the parameters to be registered has to be highlighted.
<Get FV Status> and <Set to FV> button can not be pressed during PAPP
protocol is working.
2-4-2 Editing the Protocol
The protocol of the Protocol Processor consists of several Tracks, and Track is a set of
image acquisition condition and parameter settings.
The [PAPP] window is used to edit the protocol.
The Track needs to be edited as a set of acquisition condition, then deferent Track has to be
created in case of the condition becomes different as shown below;
Track 1
Interval:
30sec
(3)
A
cquisition every 3sec
(1)
:
4 frames
(2)
Track 2
Frame Frame Frame Frame Frame Frame
3sec. 100sec.
A
cquisition every 100sec
(4)
:
2 frames
(5)
For details, see section 2-4-8,” Example of assembling a protocol”.
(1)
(2)
(4)
(5)
(3)