1-3-2-4 Configuring the filters (When using a spectral detecting system)
When the system incorporates the spectral detector unit that is composed of a 2-channel
spectral detector and 1-channel filter, it is possible to set the detection conditions more
flexibly, acquire the fluorescence spectral data and use the fluorescence isolation function.
As the upper and lower limit wavelengths can be set while acquiring images, any desired
wavelength regions can be selected easily.
To acquire image with a spectral detecting system, set the wavelength regions in the
[Spectral-Control] sub panel.
Fig. 1-18 [Spectral-Control] sub panel and [Dyes] sub panel
The settings are interlocked with those in the [Optical System Configuration] window.
Shows the graph of fluorescence
wavelength (Em_**) for each
dyes. When area inside the graph
is double clicked, a window to
change graph display method will
appear.
Set the upper limit (left) and lowe
r
limit (right) wavelengths of the
barrier filter.
Move the upper limit and lowe
r
limit wavelength of the barrier filte
r
while maintaining the width
between the upper limit and lowe
r
limit wavelengths.