Getting Started FLUOVIEW/Outline of LSM Observation Procedures
OPERATION INSTRUCTIONS
1-32
Page
1-2-7-8 Setting the Area to be Observed
When the observation targets are concentrated in a narrow area or when observation of a
specific area detail is required, the image of a limited area can be selected.
Fig. 1-9 [Pan]/[Zoom] Group Box
1-2-7-9 Setting a Lower Scan Speed
The scan speed can be decreased using the scale in the [Scan Speed] group box on the
[Acquire] panel.
In general, setting a lower scan speed allows the acquired image quality to be
improved.
However, a low scan speed also lengthens the time required for image
acquisition.
When the scan speed is decreased during fluorescence observation, the
saturation of fluorescence may darken the image of certain types of
specimens. In this case, increase the scan speed and increase the PMT
Voltage or use accumulation in scanning.
The 4 buttons represent directions, and
clicking a button moves the acquired image
area in the direction indicated by the button.
Clicking the square button on the cente
r
returns the acquired image area to the center.
Clicking a point in the scale area
to change the value on a large
scale.
Clicking the top or bottom arrow
button allows fine adjustment o
f
the value.
Dragging the square knob allows
the value to be changed directly.
Click a point inside the circle to change the
position of the acquired image area directly.
[Scan Speed] group box
Set the scan speed by clicking a
point on the scale line.
TIP
NOTE